PERATING
THE
UNIT
39
Press again the
“S”
-button
(H)
and the ramp-up time display flashes.
Now either press the “
Ÿ
”-button
(I)
or the “
ź
”-button
(J)
to set the
desired ramp-up time from 1.0 seconds to 9.0 seconds.
Press again the “
S
”-button
(H)
and the ramp-return time display flashes.
Now either press the “
Ÿ
”-button
(I)
or the “
ź
”-button
(J)
to set the
desired ramp-return time from 1.0 seconds to 9.0 seconds.
Press again the
“S”
-button
(H)
and the rest time display flashes. Now
either press the “
Ÿ
”-button
(I)
or the “
ź
”-button
(J)
to set the desired
rest time from 2 seconds to 99 seconds.
When you are done with the settings of the program mode, press the
“P”
-button
(G)
, to confirm or save the settings. The device now swit-
ches to the waiting-mode. Now press the
“S”
-button
(H)
to another pro-
gramme mode. Repeat all previous steps to set the parameters. When
you‘re done with all the settings you can start the programme by regula-
ting the intensity.
Functions of the back light
There are 3 colours of the back light that indicates the different states:
a) Back light green: Normal state
b) Back light blue: The intensity of the used channel is greater than 65mA.
With the further increase in the intensity, you must to exercise extreme
caution.
c) Back light red: The electrodes are not fixed to the skin or the device
is not connected to the electrodes, and the set intensity is greater than
10mA. The intensity is thus automatically reset to 0mA, and the back
light keeps flashing in red.
Checking the memory status
Press and hold the
“P”
-button
(G)
to reach to the saving mode. You can
now read the entire treatment time and average intensity of channel 1
or 2 with the “
Ÿ
”-button
(I)
or the “
ź
”-button
(J)
. Press the
“P”
-button
(G)
again to return to waiting mode.