ICU:CALKIT:FACT:READ
<ICU_ID>
Reads the factory characterization (factory calkit file) from the flash memory of
the related Inline Calibration Unit to the SD-Card of the Inline Calibration Control-
ler.
Caching of factory characterization data over the comparatively slow CAN bus
connection (max. data rate 1 MBit/s) is initiated automatically
●
when the respective ICU is connected to the ICC for the first time
●
if the ICC detects that the factory characterization data on the ICU have
changed (e.g. after recalibration at Rohde&Schwarz service)
Using
ICU:CALKIT:FACT:READ
it can also be triggered manually.
On the SD-card of the ICC the calkit files of a particular ICU are maintained as zip
archives in the directory
<ICU_ID>
. From there they can be downloaded via
FTP.
The factory calkit file is named
Factory.zip
, the user calkit files are named
User1.zip
,
User2.zip
, ... (see
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
Usage:
Event
Manual operation:
"Read Factory Characterization R&S ZN-Z32/Z33"
ICU:CALKIT:USER:READ
<ICU_ID>
Reads the user characterization data (i.e. all user calkit files) from the flash mem-
ory of the related Inline Calibration Unit to the SD-Card of the Inline Calibration
Controller.
On the SD-card of the ICC the calkit files of a particular ICU are maintained as zip
archives in the directory
<ICU_ID>
on page 38). From there they
can be downloaded via FTP.
The factory calkit file is named
Factory.zip
, the user calkit files are named
User1.zip
,
User2.zip
, ... (see
Remote Control Commands