![Rohde & Schwarz R&S ZN-Z3x User Manual Download Page 39](http://html.mh-extra.com/html/rohde-and-schwarz/rands-zn-z3x/rands-zn-z3x_user-manual_1477870039.webp)
Operation
R&S
®
ZN-Z3x
39
User Manual 1177.6381.02 ─ 03
4.2.2.1
ICU Properties
Commands to set/get the properties of an individual Inline Calibration Unit (ICU).
The ICUs connected to the Inline Calibration Controller can be discovered via
.
............................................................................................... 39
.................................................................................................40
...........................................................................................40
................................................................................................41
ICU:SHOW:STATE?
<ICU_ID>
Queries the state of the related Inline Calibration Unit.
The result is a composite string of the form
ICU <ICU_ID> status: <code>,
<progress>%
, e.g.
ICU 3186b6 status: 0, 0%
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
Return values:
<code>
Decimal status code as described in the table below
<progress>
Progress of a calkit transmission in percent
Usage:
Query only
Table 4-4: Inline Calibration Unit status codes
Status code
Description
0
No error
1
CAN Bus Receive Error
2
CAN Bus Transmit Error
64
24 V error
Do not connect any further Inline Calibration Unit R&S
ZN-Z32/Z33,
otherwise the calibration precision (effective system data) can no lon-
ger be guaranteed
65
13 V error
Calibration functionality degraded
Remote Control Commands