Operation
R&S
®
ZN-Z3x
43
User Manual 1177.6381.02 ─ 03
TRIGGERED
ARMED
IDLE
ICU:CAL:START
ICU:CAL:STOP
Push "Start Cal" button
Figure 4-1: "Start Cal" button state transitions
The intended use case is a calibration step requiring user interaction, for example
if a THROUGH standard has to be connected between the related Inline Calibra-
tion Units.
A remote control application should send the
poll the button state to take the appropriate actions when the user presses the
"Start Cal" button (i.e. when the
"TRIGGERED"
state is observed). Finally it
should send the
command to go back to
"IDLE"
state.
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
Return values:
<response>
0,"IDLE"
1,"ARMED"
: Waiting for the user to press the "Start Cal"
button
2,"TRIGGERED"
: "Start Cal" button was pressed in
"ARMED"
state
Usage:
Query only
4.2.2.3