Return values:
<fw_version>
Firmware version of the related ICU
Usage:
Query only
ICU:SECure:ERASe
<ICU_ID>
Sanitizes the flash memory of the related ICU (i.e. securely erases the user block
of data) or queries the sanitization state.
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
4.2.2.2
ICU Calibration
Commands to perform a calibration.
ICU:SET:PORT<no>
<ICU_ID>
Assigns port number <no> to the related Inline Calibration Unit
Suffix:
<no>
.
Port number
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
ICU:CAL:OPEN
<ICU_ID>
ICU:CAL:SHORT
<ICU_ID>
ICU:CAL:MATCH
<ICU_ID>
ICU:CAL:THROUGH
<ICU_ID>
Connects the VNA port of the related Inline Calibration Unit to its built-in OPEN,
SHORT or MATCH calibration standard or establishes the low-loss through path
between VNA and DUT port.
Note that for the initial calibration, a THROUGH standard between the DUT ports
of different Inline Calibration Units has to be connected externally.