Operation
R&S
®
ZN-Z3x
45
User Manual 1177.6381.02 ─ 03
Parameters:
<ICU_ID>
ID of a connected Inline Calibration Unit (see
Usage:
Event
Manual operation:
"Read User Characterization R&S ZN-Z32/Z33"
ICU:CALKIT:USER:WRITE
<Dest_UCkitFile>
Writes a user calkit file from the SD-Card of the Inline Calibration Controller to the
flash memory of the related Inline Calibration Unit.
If for some reason the user calkit file can't be transferred to the ICU, a "Calkit err"
(
If the ICU runs out of storage space for user characterization data, a "Memory
read/write error" (
002) is displayed at the ICU. Use
to delete existing user characterizations and free the required
storage space.
Note
●
Before uploading the calkit file to the ICC via FTP, it must be packed to a zip
archive named according to DOS 8.3 filename conventions.
●
On the SD card of the ICC, the zip archive must be placed in the related ICU's
dedicated directory
<ICU_ID>
(see
●
On the ICU the available user characterization files are maintained as
User1.zip
, ...,
User<N>.zip
. The original zip file name is lost.
Parameters:
<Dest_UCkitFile>
String identifying the related ICU and the calkit file to be
written. It is composed of the ICU_ID (see
on page 38), an @, and the name of an existing user calkit
zip file (e.g. "3186B1@new_ucal.zip").
Usage:
Event
Manual operation:
"Write User Characterization R&S ZN-Z32/Z33"
Remote Control Commands