Preparing for Use
R&S
®
NRP Series
15
Getting Started 1419.0170.02 ─ 12
Setup
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (−70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 4-1: Setup with an R&S NRP
‑
ZKU cable
1 = Signal source
2 = R&S NRP power sensor
3 = Host interface connector
4 = R&S NRP
‑
ZKU cable
5 = USB connector
6 = Computer with installed VISA driver or R&S NRP Toolkit
Incorrectly connecting/disconnecting the R&S NRP power sensors can
damage the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Chapter 4, "Preparing for Use"
1. Connect the cables as shown in
a) Connect the R&S NRP
‑
ZKU cable to the power sensor.
"To connect a cable to the host interface of the power sensor"
b) Connect the R&S NRP
‑
ZKU cable to the computer.
c) Connect the power sensor to the signal source.
2. On the computer, start a software application to view the measurement
results.
See
Chapter 6, "Operating Concepts"
4.6.1.2
R&S NRP
‑
Z5 Sensor Hub Setup
The R&S NRP
‑
Z5 sensor hub (high-speed USB 2.0) can host up to four R&S
NRP power sensors and provides simultaneous external triggering to all connec-
ted sensors.
Connecting to a Controlling Host