Operating Concepts
R&S
®
NRP Series
38
Getting Started 1419.0170.02 ─ 12
Incorrectly connecting/disconnecting the R&S NRP power sensors can
damage the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in
Chapter 4, "Preparing for Use"
Starting a measurement
1. Connect the power sensor to the computer as shown in
For a detailed description, refer to
Chapter 4.6.1.1, "Simple USB Connection"
2. Start the R&S NRPV.
3. Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measure-
ment signal during zeroing causes an error.
a) Switch off the measurement signal.
b) Select "Zero > Select > A" (channel short name).
Zeroing takes several seconds. During zeroing, a message shows the pro-
gress. After completion, the message reports either success or an error
("Success" / "Failed").
4. Switch on the test signal of the signal source.
5. To start a continuous measurement, select "Measure > Continuous".
The "Continuous" measurement window appears. It shows the measurement
results numerically, and the control panel for accessing further dialogs with
parameters for measurement, evaluation and display.
R&S NRPV