Operating Concepts
R&S
®
NRP Series
45
Getting Started 1419.0170.02 ─ 12
b) Connect the R&S NRP
‑
ZK6 cable to a sensor input connector of the R&S
NRP2.
c) Connect the [RF] connector of the power sensor to the signal source.
2. Preset the R&S NRP2.
a) Press the [(PRE)SET] hardkey.
The "File" menu appears.
b) Press the [(PRE)SET] hardkey again or press the "Preset" softkey.
All parameters are set to their defaults, even when in inactive operating
modes.
3. Execute zeroing:
Note:
Turn off all measurement signals before zeroing. An active measure-
ment signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [ZERO] hardkey of the R&S NRP2.
The "Zero" dialog box is displayed.
c) Press the [ZERO] hardkey again to perform zeroing of all connected sen-
sor channels ("Zero (All)") or press the appropriate softkey to select a spe-
cific sensor for zeroing.
4. Press the [FREQ] hardkey and enter the carrier frequency of the applied sig-
nal if the specified measurement accuracy is to be reached.
5. Switch on the signal source.
The result window indicates the result (in dBm) obtained with sensor A.
R&S NRP2