![Rohde & Schwarz FSMR series Getting Started Download Page 78](http://html.mh-extra.com/html/rohde-and-schwarz/fsmr-series/fsmr-series_getting-started_1477836078.webp)
R&S FSMR
The
Screen
1166.3363.12
3.5
E-1
Status information
#SMPL
UNCAL
The status information on the lef t side of the diagram hint at irregularity
(e.g. UNCAL)
"#SMPL" indicates that the relation Span / RBW is higher than 125 while
the RMS detector is activated. In this case, a stable signal evaluation is
no longer possible due to an ins ufficient number of A/D c onverter
samples.
⇒
reduce span or increase RBW
"UNCAL" is indicated under the following circumstances:
•
correction data are switched off (menu CAL, CAL CORR OFF).
⇒
switch on CAL CORR ON or PRESET
•
no valid correction data. T his may occur after a c old start of the
instrument following a firmware update.
⇒
record correction data
•
Sweep time too s hort for current instrument settings (span,
resolution bandwidth, video bandwidth).
⇒
increase sweep time
OVLD / IFOVL
LOUNL /
EXREF
OVEN
OVLD is indicated when the input mixer is overloaded.
⇒
Increase input attenuation
IFOVL is indicated when overload occurs in the IF signal path after the
input mixer.
⇒
Increase reference level
LOUNL is indicated when an error occurs in the frequency processing of
the instrument.
EXREF is indicated when the analy zer is configured for use of an
external reference signal, but no r eference signal is detected at the
corresponding input.
OVEN is indicated when the c rystal oscillator (option R&S FSMR-B4)
has not yet reached its operating temperature. This indication vanishes
after a few minutes after switch on.
Trace info:
Every active measurement curve (trace
≠
BLANK) is allocated trace
information of two or three lines at the left of the diagram. The trace
information has the same colour as the measurement curve.
The information on the currently selected trace is displayed in inverse
video (see
TRACE - SELECT TRACE softkey
).
Summary of Contents for FSMR series
Page 1: ...R S FSMR Measuring Receiver Getting Started Test Measurement Getting Started 1313 9698 02 03 ...
Page 7: ......
Page 9: ......
Page 11: ...Contents Preparing for Operation R S FSMR 1313 9723 12 I 1 2 E 1 Fig 1 1 Front View ...
Page 13: ...Front View R S FSMR 1313 9723 12 1 2 E 2 Fig 1 1 Front View ...
Page 15: ...Front View R S FSMR 1313 9723 12 1 4 E 2 Fig 1 1 Front View ...
Page 17: ...Front View R S FSMR 1313 9723 12 1 6 E 2 Fig 1 1 Front View ...
Page 19: ...Front View R S FSMR 1313 9723 12 1 8 E 2 Fig 1 1 Front View ...
Page 51: ......
Page 71: ......
Page 102: ...R S FSMR Menu Overview 1166 3363 12 3 29 E 3 AMPT Key ...
Page 107: ...Menu Overview R S FSMR 1166 3363 12 3 34 E 3 DEMOD Key ...
Page 108: ...R S FSMR Menu Overview 1166 3363 12 3 35 E 3 AUDIO Key ...
Page 111: ...Menu Overview R S FSMR 1166 3363 12 3 38 E 3 AMPT Key ...
Page 112: ...R S FSMR Menu Overview 1166 3363 12 3 39 E 3 MEAS Key ...
Page 113: ...Menu Overview R S FSMR 1166 3363 12 3 40 E 3 ...
Page 114: ...R S FSMR Menu Overview 1166 3363 12 3 41 E 3 BW Key ...
Page 116: ...R S FSMR Menu Overview 1166 3363 12 3 43 E 3 Menu Overview another Keys MKR Key ...
Page 120: ...R S FSMR Menu Overview 1166 3363 12 3 47 E 3 TRACE Key ...
Page 123: ...Menu Overview R S FSMR 1166 3363 12 3 50 E 3 FILE Key ...
Page 125: ...Menu Overview R S FSMR 1166 3363 12 3 52 E 3 SETUP Key ...
Page 127: ...Menu Overview R S FSMR 1166 3363 12 3 54 E 3 LOCAL Menu LOCAL ...
Page 128: ...R S FSMR Menu Overview 1166 3363 12 3 55 E 3 Menu Overview Ext Generator Control ...