Connecting the Instrument to the AC Supply
The R&S FSMR is equipped with an AC voltage selection feature and will automatically adapt itself to
the applied AC voltage (range: 100 to 240 V AC, 40 to 400 Hz). External voltage selection or adaptation
of the fuses are not necessary. The AC power connector is located on the rear panel (see below).
o I
Power connector
Connect the instrument to the AC power source using the AC
power cable delivered with the instrument.
As the instrument is designed according to the regulations for
safety class EN61010, it must be connected to a power outlet
with earthing contact.
Switching the Instrument On/Off
Do not power down during booting.
Such a switch-off may lead to corruption of the hard disk files
AC power switch on the rear panel
Power switch
I 0
Power connector
Power switch
Position I = ON
In the I position, the instrument is in standby mode or in
operation, depending on the position of the ON/STANDBY
key at the front of the instrument
.
Note:
The AC power switc
h
may remain ON continuously.
Switching to OFF is only required when the instrument
must be completely removed from the AC power source.
Position O = OFF
The 0 position implies an all-pole disconnection of the
instrument from the AC power source.
ON/STANDBY switch on the front panel
ON STANDBY
Standby switch
The ON/STANDBY switch activates two different operating
modes indicated by coloured LEDs:
Operation ON -
ON/STANDBY is depressed
The green LED (ON) is illuminated. The instrument is
ready for operation. All modules within the instrument are
supplied with power.
Shock hazard
In standby mode, the AC power voltage is present within the instrument
STANDBY -
ON/STANDBY switch is not pressed.
The yellow LED (STANDBY) is illuminated. Only the
power supply is supplied with power and the quartz oven
is maintained at normal operating temperature.
Summary of Contents for FSMR series
Page 1: ...R S FSMR Measuring Receiver Getting Started Test Measurement Getting Started 1313 9698 02 03 ...
Page 7: ......
Page 9: ......
Page 11: ...Contents Preparing for Operation R S FSMR 1313 9723 12 I 1 2 E 1 Fig 1 1 Front View ...
Page 13: ...Front View R S FSMR 1313 9723 12 1 2 E 2 Fig 1 1 Front View ...
Page 15: ...Front View R S FSMR 1313 9723 12 1 4 E 2 Fig 1 1 Front View ...
Page 17: ...Front View R S FSMR 1313 9723 12 1 6 E 2 Fig 1 1 Front View ...
Page 19: ...Front View R S FSMR 1313 9723 12 1 8 E 2 Fig 1 1 Front View ...
Page 51: ......
Page 71: ......
Page 102: ...R S FSMR Menu Overview 1166 3363 12 3 29 E 3 AMPT Key ...
Page 107: ...Menu Overview R S FSMR 1166 3363 12 3 34 E 3 DEMOD Key ...
Page 108: ...R S FSMR Menu Overview 1166 3363 12 3 35 E 3 AUDIO Key ...
Page 111: ...Menu Overview R S FSMR 1166 3363 12 3 38 E 3 AMPT Key ...
Page 112: ...R S FSMR Menu Overview 1166 3363 12 3 39 E 3 MEAS Key ...
Page 113: ...Menu Overview R S FSMR 1166 3363 12 3 40 E 3 ...
Page 114: ...R S FSMR Menu Overview 1166 3363 12 3 41 E 3 BW Key ...
Page 116: ...R S FSMR Menu Overview 1166 3363 12 3 43 E 3 Menu Overview another Keys MKR Key ...
Page 120: ...R S FSMR Menu Overview 1166 3363 12 3 47 E 3 TRACE Key ...
Page 123: ...Menu Overview R S FSMR 1166 3363 12 3 50 E 3 FILE Key ...
Page 125: ...Menu Overview R S FSMR 1166 3363 12 3 52 E 3 SETUP Key ...
Page 127: ...Menu Overview R S FSMR 1166 3363 12 3 54 E 3 LOCAL Menu LOCAL ...
Page 128: ...R S FSMR Menu Overview 1166 3363 12 3 55 E 3 Menu Overview Ext Generator Control ...