![Rohde & Schwarz FSMR series Getting Started Download Page 66](http://html.mh-extra.com/html/rohde-and-schwarz/fsmr-series/fsmr-series_getting-started_1477836066.webp)
R&S FSMR
Measurement Examples
1166.3311.12
2.15
E-2
DEMOD
RELATIVE
+- PEAK/2
2. Relative Audio measurements – audio frequency response
Press the
DEMOD
hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the
RELATIVE
softkey.
The RELATIVE menu is displayed.In the relative menu, only the
detectors which have been switched on in the DETECTOR menu
are available for relative measurements.
In the RELATIVE menu press the
+- PEAK/2
softkey.
The actual measurement value of the detector will be saved as a
reference value and the result display will change to a relative
reading (indicated with
K
0.00 %).
Change the modulation frequency to 50 kHz
The following settings are used on the generator:
1. Select the AM Modulation, 50 kHz modulation frequency.
RELATIVE
RELATIVE
DB %
Relative display – Change the Unit to dB
Press the
DEMOD
hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the
RELATIVE
softkey.
The RELATIVE menu is displayed.
In the RELATIVE menu press the
+- PEAK/2
softkey.
The actual result display will change from a relative reading
indicated in % to a dB reading.
Fig.2-11
Relative measurement in the demod mode
Summary of Contents for FSMR series
Page 1: ...R S FSMR Measuring Receiver Getting Started Test Measurement Getting Started 1313 9698 02 03 ...
Page 7: ......
Page 9: ......
Page 11: ...Contents Preparing for Operation R S FSMR 1313 9723 12 I 1 2 E 1 Fig 1 1 Front View ...
Page 13: ...Front View R S FSMR 1313 9723 12 1 2 E 2 Fig 1 1 Front View ...
Page 15: ...Front View R S FSMR 1313 9723 12 1 4 E 2 Fig 1 1 Front View ...
Page 17: ...Front View R S FSMR 1313 9723 12 1 6 E 2 Fig 1 1 Front View ...
Page 19: ...Front View R S FSMR 1313 9723 12 1 8 E 2 Fig 1 1 Front View ...
Page 51: ......
Page 71: ......
Page 102: ...R S FSMR Menu Overview 1166 3363 12 3 29 E 3 AMPT Key ...
Page 107: ...Menu Overview R S FSMR 1166 3363 12 3 34 E 3 DEMOD Key ...
Page 108: ...R S FSMR Menu Overview 1166 3363 12 3 35 E 3 AUDIO Key ...
Page 111: ...Menu Overview R S FSMR 1166 3363 12 3 38 E 3 AMPT Key ...
Page 112: ...R S FSMR Menu Overview 1166 3363 12 3 39 E 3 MEAS Key ...
Page 113: ...Menu Overview R S FSMR 1166 3363 12 3 40 E 3 ...
Page 114: ...R S FSMR Menu Overview 1166 3363 12 3 41 E 3 BW Key ...
Page 116: ...R S FSMR Menu Overview 1166 3363 12 3 43 E 3 Menu Overview another Keys MKR Key ...
Page 120: ...R S FSMR Menu Overview 1166 3363 12 3 47 E 3 TRACE Key ...
Page 123: ...Menu Overview R S FSMR 1166 3363 12 3 50 E 3 FILE Key ...
Page 125: ...Menu Overview R S FSMR 1166 3363 12 3 52 E 3 SETUP Key ...
Page 127: ...Menu Overview R S FSMR 1166 3363 12 3 54 E 3 LOCAL Menu LOCAL ...
Page 128: ...R S FSMR Menu Overview 1166 3363 12 3 55 E 3 Menu Overview Ext Generator Control ...