Measurement Examples
R&S FSMR
1166.3311.12
2.16
E-2
DEMOD
DETECTOR
SELECTION
DISTORTION
& SINAD
3. Measure the harmonic distortion - Choosing the detector
Press the
DEMOD
hotkey.
In the DEMOD menu press the
DETECTOR SELECTION
softkey.
The DETECTOR SELECTION menu is displayed
The instrument is equipped with a wide range of detectors for
measuring modulation signals. The peak detectors are a good
choice for capturing the highest positive or negative peak within the
measurement time, while the RMS and Average detectors are the
best choice for measuring noise and residual modulation. The THD
& SINAD detector measure the distortion of the demodulated audio
signal. All detectors can be used in parallel.
In the DETECTOR SELECTION menu press the
DISTORTION &
SINAD
softkey.
AMPT
UNIT
DIST UNIT
%
DB
Total Harmonic Distortion – Change the Unit to %
Press the
AMPT
hardkey on the frontpanel.
The amplitude settings menu is displayed.
In the AMPLITUDE menu press the
UNIT
softkey.
The measuring receiver UNIT menu is displayed.
In the UNIT menu press the
DIST UNIT % / DB
softkey.
The actual result display for the THD measurement will change from
a “DB” reading to a “%” reading.
Fig.2-12
THD measurement in UNIT %
Summary of Contents for FSMR series
Page 1: ...R S FSMR Measuring Receiver Getting Started Test Measurement Getting Started 1313 9698 02 03 ...
Page 7: ......
Page 9: ......
Page 11: ...Contents Preparing for Operation R S FSMR 1313 9723 12 I 1 2 E 1 Fig 1 1 Front View ...
Page 13: ...Front View R S FSMR 1313 9723 12 1 2 E 2 Fig 1 1 Front View ...
Page 15: ...Front View R S FSMR 1313 9723 12 1 4 E 2 Fig 1 1 Front View ...
Page 17: ...Front View R S FSMR 1313 9723 12 1 6 E 2 Fig 1 1 Front View ...
Page 19: ...Front View R S FSMR 1313 9723 12 1 8 E 2 Fig 1 1 Front View ...
Page 51: ......
Page 71: ......
Page 102: ...R S FSMR Menu Overview 1166 3363 12 3 29 E 3 AMPT Key ...
Page 107: ...Menu Overview R S FSMR 1166 3363 12 3 34 E 3 DEMOD Key ...
Page 108: ...R S FSMR Menu Overview 1166 3363 12 3 35 E 3 AUDIO Key ...
Page 111: ...Menu Overview R S FSMR 1166 3363 12 3 38 E 3 AMPT Key ...
Page 112: ...R S FSMR Menu Overview 1166 3363 12 3 39 E 3 MEAS Key ...
Page 113: ...Menu Overview R S FSMR 1166 3363 12 3 40 E 3 ...
Page 114: ...R S FSMR Menu Overview 1166 3363 12 3 41 E 3 BW Key ...
Page 116: ...R S FSMR Menu Overview 1166 3363 12 3 43 E 3 Menu Overview another Keys MKR Key ...
Page 120: ...R S FSMR Menu Overview 1166 3363 12 3 47 E 3 TRACE Key ...
Page 123: ...Menu Overview R S FSMR 1166 3363 12 3 50 E 3 FILE Key ...
Page 125: ...Menu Overview R S FSMR 1166 3363 12 3 52 E 3 SETUP Key ...
Page 127: ...Menu Overview R S FSMR 1166 3363 12 3 54 E 3 LOCAL Menu LOCAL ...
Page 128: ...R S FSMR Menu Overview 1166 3363 12 3 55 E 3 Menu Overview Ext Generator Control ...