Preparing for use
R&S
®
SMM100A
24
Getting Started 1179.1335.02 ─ 04
Table 3-1: Connector types and torque limits
Type
Torque limit
Nut opening
lb-Inch
Nm
Inch
mm
N
13.3
1.5
3/4
20
SMA
5
0.56
5/16
8
3.5 mm
8
0.9
5/16
8
2.92 mm
8
0.9
5/16
8
2.4 mm
8
0.9
5/16
8
1.85 mm
8
0.9
5/16
8
1.0 mm
3
0.34
0.236
6
To prevent RF output switch-off
►
NOTICE!
If you set a too high output level without a load connected to the
instrument, the reverse power can exceed a limit forcing the R&S
SMM100A
to switch off the RF output.
Connect a load with sufficient return loss as given in the data sheet.
3.10
Connecting to LO In/Out
"LO In/Out" connectors are Subminiature Version A (SMA) connectors.
The connector is located on the
"To connect to screwable connectors"
See also chapter "Local Oscillator Coupling" in the user manual.
3.11
Connecting to Ref In/Ref Out
The connector is located on the
Connecting to Ref In/Ref Out
Summary of Contents for 1440.8002.02
Page 1: ...R S SMM100A Vector Signal Generator Getting Started 1179133502 Version 04 Ý S2 ...
Page 6: ...Contents R S SMM100A 6 Getting Started 1179 1335 02 04 ...
Page 14: ...Key features R S SMM100A 14 Getting Started 1179 1335 02 04 ...
Page 28: ...Preparing for use R S SMM100A 28 Getting Started 1179 1335 02 04 Switching on or off ...
Page 44: ...Instrument tour R S SMM100A 44 Getting Started 1179 1335 02 04 Rear panel tour ...
Page 90: ...Instrument control R S SMM100A 90 Getting Started 1179 1335 02 04 Remote operation over VNC ...
Page 92: ...Contacting customer support R S SMM100A 92 Getting Started 1179 1335 02 04 ...