Preparing for use
R&S
®
SMM100A
23
Getting Started 1179.1335.02 ─ 04
3. If the R&S
SMM100A is switched on, deactivate the RF output, before con-
necting an RF cable to the RF connector.
In the block diagram, select the block "RF" > "RF Level" > "RF ON > Off".
4. Use a high-quality RF cable that matches the RF connector type.
"Cable selection and electromagnetic interference (EMI)"
To connect to non-screwable connectors (BNC)
► To connect the RF cable with the RF connector, proceed as follows:
a) Carefully align the connector of the cable and the RF connector along a
common axis.
b) Mate the connectors along the common axis until the male pin of the con-
nector of the cable engages with the female socket of the RF connector.
To connect to screwable connectors
1. Use a high-quality cable that matches the connector type.
"Cable selection and electromagnetic interference (EMI)"
2.
NOTICE!
Risk of instrument damage and connector damage. Excessive tight-
ening can damage the cables and the connectors. However, if you do not
tighten the connectors enough, the measurement results can be inaccurate.
To connect the cable with the connector, proceed as follows:
a) Carefully align the connector of the cable and the connector along a com-
mon axis.
b) Mate the connectors along the common axis until the male pin of the inner
connector engages with the female socket of the outer connector.
c) Turn the nut of the outer connector until the connectors are firmly coupled.
d) Torque the nut to the specified limit using a calibrated torque wrench. Hold
the opposite connector part stationary with a spanner.
For torque limits of the most relevant connector types, see
For more information, see chapter "Handling" of the application note
If your instrument is equipped with a test port adapter, see the application note
.
The connector types listed in this table represent the common connectors provi-
ded by Rohde
&
Schwarz. It is considered as general information and therefore
can contain connector types that do not apply to your instrument.
See
Connecting to RF
Summary of Contents for 1440.8002.02
Page 1: ...R S SMM100A Vector Signal Generator Getting Started 1179133502 Version 04 Ý S2 ...
Page 6: ...Contents R S SMM100A 6 Getting Started 1179 1335 02 04 ...
Page 14: ...Key features R S SMM100A 14 Getting Started 1179 1335 02 04 ...
Page 28: ...Preparing for use R S SMM100A 28 Getting Started 1179 1335 02 04 Switching on or off ...
Page 44: ...Instrument tour R S SMM100A 44 Getting Started 1179 1335 02 04 Rear panel tour ...
Page 90: ...Instrument control R S SMM100A 90 Getting Started 1179 1335 02 04 Remote operation over VNC ...
Page 92: ...Contacting customer support R S SMM100A 92 Getting Started 1179 1335 02 04 ...