3.5
Considerations for test setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, especially for the following connector types:
– BNC
Double-shielded BNC cables.
How to:
"To connect to non-screwable connectors (BNC)"
– SMA
Double-shielded SMA cables.
How to:
"To connect to Ref In/Ref Out (reference = 1
– USB
Double-shielded USB cables.
How to:
Chapter 3.8, "Connecting USB devices"
See also chapter "Troubleshooting and Error Messages" in the user man-
ual.
– LAN
At least CAT6 STP cables.
How to:
Chapter 3.7, "Connecting to LAN"
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
●
Use cables of the same type and equal length for connection to the I/Q and
I/Q Bar interfaces of the instrument.
●
Use the cable R&S
SMU-Z6 for connection to the Dig I/Q interfaces of the
instrument. The cable is available under material number 1415.0201.02.
●
Use the cable R&S
DIGIQ-HS for connection to the HS Dig I/Q interfaces of
the instrument. The cable is available under material number 3641.2948.03.
How to:
Chapter 3.12, "Connecting to HS Dig I/Q"
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels
within the specified ranges to avoid damage to the R&S
SMM100A and connec-
ted devices.
Considerations for test setup
Summary of Contents for 1440.8002.02
Page 1: ...R S SMM100A Vector Signal Generator Getting Started 1179133502 Version 04 Ý S2 ...
Page 6: ...Contents R S SMM100A 6 Getting Started 1179 1335 02 04 ...
Page 14: ...Key features R S SMM100A 14 Getting Started 1179 1335 02 04 ...
Page 28: ...Preparing for use R S SMM100A 28 Getting Started 1179 1335 02 04 Switching on or off ...
Page 44: ...Instrument tour R S SMM100A 44 Getting Started 1179 1335 02 04 Rear panel tour ...
Page 90: ...Instrument control R S SMM100A 90 Getting Started 1179 1335 02 04 Remote operation over VNC ...
Page 92: ...Contacting customer support R S SMM100A 92 Getting Started 1179 1335 02 04 ...