5-1
■
5. TROUBLESHOOTING
Check the following first when you encounter any problems.
If the problem is not solved even after checking the applicable items listed below, contact
your local distributor to request inspection and/or repair.
■
Do not remove the cover of the instrument. Otherwise you may get electric
shock.
■
Do not take any actions other than those specified below.
5.1 Common items
The instrument does not start when the power switch is
turned on.
Problem with the power plug
Check that the power plug is firmly connected to the
outlet. Check that there are no flaws in the power cord,
such as cracks or tears.
Problem with the power outlet
Check that power is supplied to the outlet to which the
power cord is connected.
Faulty fuse
Check that fuses are not blown. If blown, replace the
fuse (refer to “4.5.1 Fuses”).
If the new fuse blows
again, the instrument may be faulty. Contact your local
distributor to request inspection and/or repair.
Nothing appears on the monitor screen.
The auto power off function, which automatically turns
off the screen when the instrument is not operated for a
specified time, has been activated.
Touch the monitor screen.
The maintenance switch on the side is in the P (up)
position.
Turn off the power, return the maintenance switch to
the down position, and then turn on the power.