3-26
1
)
Specify the range.
- Touch the range selection button (4) to activate it. Touch two opposing
points of the area (rectangle) subject to calculation on the image of
the endothelium.
-
When “100 μm” is set to “
L-
count range” of the system setup items,
touch the center of the 100-
μ
m-
square frame. Refer to “3.7.3
Application
”
for the system setup.
- The area of the specified range is shown in the Area field (3).
-
When “L
-
count Assist” in the system setup is set to ON, cells in the
selected range are simply extracted and dots appear on the cells.
2
)
Place a dot on a cell.
-
Touch the cell selection button (5) to activate it. Specify the center of
all cells in the specified range using the touch pen.
-
An orange dot appears at the touched point.
-
The number of current dots appears in the
“Count”
field (3).
-
To delete a dot, touch the eraser button (7) to activate it and then
touch the dot to be deleted.
-
Touching and holding the all delete button (6) deletes all dots
displayed.
-
Touch the undo/redo buttons (8) to undo and/or redo the operation
to specify dots.
3)
Touch the “Calc. (calculation)”
button (9) to close the Point Edit screen.
-
The calculation results are displayed on the analysis screen.
-
Touching the
“Cancel”
button (10) will discard the edited point
information, and return you to the analysis screen.
■
The range line in the L-count method is shown in orange (solid line) and light
blue (dotted line).
■
To specify cells, point to cells of which outlines are crossed with an orange
solid line. Do not point to cells if their outlines are crossed with a light blue
dotted line.
(Fig. 4)