3-45
■
3.7.3 Application
Make settings related to display and analysis. Select the “Application” tab.
(Fig. 1)
(1) Analysis Method
Selects the switching method for analysis method.
Set the function for the lower button 1 on the analysis screen (Fig. 2).
(Fig. 2)
(2) Fixation
Sets whether to show the image capturing position on the fixation light
position display of the analysis screen.
a) CCT
Touch the “Setting” button (1) of “CCT” to display the System Setup
screen
(Fig. 2).
(Fig. 1)
(Fig. 2)
(2) CCT (US) Display
Sets whether to display the ultrasound correction value of the central
corneal thickness.
The default setting is OFF.
(3) CCT (US) Offset
Sets the offset for ultrasound correction of central corneal thickness.
Default setting
:
13
μ
m
Setting range
:
0
μ
m - 30
μ
m (in increments of 1
μ
m)
This item is not displayed when (2)
“CCT (US) Display” is set to OFF.
(1)
(1)
(2)
(3)
(2)
(Button 1)