3-46
b) Auto Analysis
Touch the
“Setting” button (1) of “Auto Analysis” to open the System Setup
screen (Fig. 2).
(Fig. 1) (Fig. 2)
(2) Analysis result
Sets images (Photo, Trace, Area or Apex) to be shown immediately after
analysis is completed.
(3) Reliability
Sets whether to display the reliability mark on the screen.
(4) Histogram Range
Sets the range displayed by a histogram of area-specific distribution.
Default setting
:
100
μ
m
2
Setting range
:
1 - 999
μ
m
2
(in increments of 1
μ
m
2
)
(5) Dark Area Analysis
Turns the Dark Area detection function on and off.
Dark Areas are not detected when set to OFF.
(6) Dark Area as Cell
Sets whether to regard detected Dark Areas as endothelial cells and
include them in analysis of endothelial cells. This item is not displayed
when (5)
“Dark Area Analysis” is set to OFF.
(7) Analysis Area
Sets how to specify the range using the analysis range selection button
of the auto analysis/edit trace line function.
Variable
:
Set the analysis range by specifying two opposing
points.
100 μm
:
Set the analysis range to 100 μm
2
regarding a point
specified on the image of endothelium as the center.
(1)
(2)
(3)
(4)
(5)
(6)
(7)