3.3.12 Displaying reliability
The reliability level for automatic analysis is automatically
determined and displayed.
Select “ON” for the “Reliability” setting on the Setup screen
to display the reliability level. (Refer to “3.6.2 Function.”)
The reliability is categorized into 3 levels and indicated by
the mark corresponding to each level as listed below.
Reliability mark : Description
No mark : The reliability for automatic analysis is high
because cells are clearly captured over a
wide range and cell outlines are extracted
precisely by automatic analysis.
e
: Because the range of captured cells is
slightly less or visibility of captured images
is slightly poor, some cell outlines are not
extracted correctly by automatic analysis
and the resultant reliability is slightly lower.
E
: Because the range of captured cells is small
and visibility of captured images is poor,
cell outlines are not extracted correctly
by automatic analysis and the resultant
reliability is low.
When “e” or “E” is displayed, perform any of the following
actions according to the captured image condition.
n
Touch the “Select” button to return to the captured
image selection screen and perform automatic
analysis again using another image. (Refer to “3.3.7
Captured image selection screen.”)
n
Correct analysis calculation results using the
Select/Cancel button. (Refer to “Select/Cancel
buttons” of “3.3.8 Analysis result screen.”)
n
Edit cell outlines using the “Edit/Edit R/Edit L”
button. (Refer to “3.3.9 Editing outline extraction
result.”)
n
Perform analysis using the L-count function. (Refer
to “3.3.10 Analytical calculation using the L-count
function.”
n
Capture another image or perform automatic
analysis again.
3-28
Summary of Contents for REM 3000
Page 2: ......
Page 10: ...This page is intentionally blank iii 4 ...
Page 26: ...2 7 ...
Page 29: ...2 9 ...
Page 31: ...2 10 ...
Page 39: ...3 3 ...
Page 46: ...3 9 ...
Page 56: ...3 18 ...
Page 64: ...3 23 ...
Page 66: ... 15 6 Fig 2 3 24 8 9 7 16 Redo button 9 ...
Page 84: ...This page is intentionally blank 3 42 ...
Page 86: ...This page is intentionally blank 4 2 ...
Page 89: ...5 2 ...
Page 99: ...This page is intentionally blank 7 2 ...
Page 107: ...W Warranty 5 1 9 2 ...
Page 108: ......
Page 109: ...20141218 ...