4.5.13 Maddox test (vertical phoria)
Use: To detect vertical phoria
Chart: Fixation
Auxiliary lens: Rotary prism to the right eye, vertical maddox rod to the left eye
Ideal appearance:
Right eye
Left eye
Binocular ideal
1. Present the fixation chart.
Press
The horizontal maddox rod is set in the right measuring window.
The base in/out prism powers can be entered.
2. Press again.
The maddox rod is removed from the right measuring window and the vertical maddox
rod is placed in the left measuring window.
The base up/down prism powers can be entered.
3.
Ask the patient “Is the white spot above or below the red bar? Or, it is overlaid
on the
red b
ar?”
When the spot is shifted upward or downward, correct it as follows:
Appearance of Chart
Phoria
Correction
1) The white spot is above
the red bar.
Left eye
hyperphoria
Press
Turn the dial clockwise to add the
BU prism power until the spot is overlaid on the
red bar.
2) The white spot is below
the red bar.
Right eye
hyperphoria
Press
Turn the dial counterclockwise to
add the BD prism power until the spot is
overlaid on the red bar.
Use
or
instead of the dial for fine adjustment.
Summary of Contents for Phoromat 2000
Page 45: ...Phoromat 2000 Operation Manual 37 3 2...
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 207: ......
Page 215: ...11 Specifications 12 Components 13 EMC ELECTROMAGNETIC COMPATIBILITY...
Page 229: ...9 Types of Chart 9 1 A Type...