Phoromat 2000 Operation Manual
- 28 -
13. Refine the spherical power with the red-
green chart.
1) Press
The SPH +0.50 D lens is automatically added to
fog the vision. The red-green chart is presented.
2) Reduce
fogging
gradually
until
the
sharpness of the letters on the red and
green sides appears equal.
Letters on the red side are sharper
Turn the dial clockwise by increments of one.
Letters on the green side are sharper
Turn the dial counterclockwise by increments of
one
14. Refine the spherical power as high as possible to achieve the best possible
visual acuity.
1) Press
The isolated 20/20 line is presented.
2) Check the best possible visual acuity with
or .
3) Refine the spherical power as high as
possible to achieve the best possible visual
acuity.
Turn the dial counterclockwise to add SPH +0.25D
When the chart appears blur, turn the dial clock-wise by increments of one.
The spherical power for the right eye is perfectly refined.
15. Follow the same procedures as Steps 9 to
14 for the left eye.
The spherical power for the left eye is perfectly
refined.
16. Press
Summary of Contents for Phoromat 2000
Page 45: ...Phoromat 2000 Operation Manual 37 3 2...
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 207: ......
Page 215: ...11 Specifications 12 Components 13 EMC ELECTROMAGNETIC COMPATIBILITY...
Page 229: ...9 Types of Chart 9 1 A Type...