Phoromat 2000 Operation Manual
- 72 -
4.5.9 Schober test
Use: To correct heterophoria
Chart: Schober
Applicable model: Rodachart 420
Auxiliary lens: Red filter on the right eye, green filter on the left eye, binocular rotary prism
Ideal appearance:
Right eye
Left eye
Binocular ideal
[Procedure example]
1. Present the Schober chart.
Press
The red filter is placed in the right measuring window and the green filter is placed in
the left measuring window.
The base in/out prism powers can be entered.
When the Schober chart button is pressed again, the base up/down prism powers can
be entered.
2.
Ask the patient, “Can you see a green circle and red cross?”
Yes → Go to the next step.
No → Impossible to continue the test.
3. Ask the
patient, “Is the cross in the center of the circle?”
Yes
Orthophoria
No
Heterophoria
–
Go to the next step.
4.
Ask the patient, “Is the cross shifted to the left or right?”
Right
Esophoria
Left
Exophoria
The cross is in the center. → No horizontal
phoria
5.
Ask the patient, “Is the cross shifted upward or downward?”
Upward
Left eye hyperphoria
Downward
Right eye hyperphoria
The cross is in the center.
No vertical phoria
Summary of Contents for Phoromat 2000
Page 45: ...Phoromat 2000 Operation Manual 37 3 2...
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 207: ......
Page 215: ...11 Specifications 12 Components 13 EMC ELECTROMAGNETIC COMPATIBILITY...
Page 229: ...9 Types of Chart 9 1 A Type...