Phoromat 2000 Operation Manual
- 45 -
• Cross cylinder test
• Fog amount
• Data to be entered (Unaided to Final)
• Auxiliary lens
es
• With or without beep sounds
• Whether the near point lamp lights up or goes out
NOTE
- It is impossible to select the cross cylinder setting between Flipped and Auto. Set
the lens type with the Cross cylinder (XC) test parameter.
- It is possible to add the fog function with both eyes open. Set the fog according to
“3.4 Fog Function with Both Eyes Open”
. The setting is programmed the same as
other auxiliary lenses.
6. When the steps on the screen are programmed, go to the next page.
Press
on the middle of right of the screen.
When all steps are programmed, press [Complete].
7. Repeat Steps 4 and 6 until the desired charts are programmed in the
presentation order.
It is possible to store up to 30 steps.
8. Exit from programming mode.
Press [End] until the screen returns to the normal measurement screen.
3.9.2 Operating programs
The charts are presented in the programmed order.
1. Select the program to be used with
+ .
2. Start the program.
Press
The first chart in the program is presented and the refractor head is set as
programmed.
3. Present the next chart.
Press
Each time the key is pressed, the chart is presented in the programmed
order and the refractor is set as programmed.
+
Returns to the previous chart
Summary of Contents for Phoromat 2000
Page 45: ...Phoromat 2000 Operation Manual 37 3 2...
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......
Page 114: ......
Page 115: ......
Page 116: ......
Page 207: ......
Page 215: ...11 Specifications 12 Components 13 EMC ELECTROMAGNETIC COMPATIBILITY...
Page 229: ...9 Types of Chart 9 1 A Type...