
Roche Diagnostics
Operator’s Manual · Version 1.0
C-19
URISYS 2400
5 Daily operation
Replace an expired URISYS 2400 cassette
Replace an expired URISYS 2400 cassette
When a URISYS 2400 cassette has exceeded its 14-day on-board stability, and
the cassette is not yet empty, it must be exchanged with a new one.
a
To replace an expired URISYS 2400 cassette
1
Press <Menu> on the [Routine Monitor] screen. The [Menu] screen is
displayed.
2
Press <Maintenance>. The [Maintenance] screen is displayed.
3
Press <Change Cassette>. The [Change Cassette] screen is displayed.
4
Press <Remove>. A confirmation window appears.
5
Press <Execute>. The cassette base rises to the removal position. A “Do
not open the door” message is displayed, followed by an “Exchange
cassette and press <Yes> key” message.
6
Open the cassette compartment door.
7
Remove the cassette by pulling the handle on the left side of the cassette
and take out the cassette.
8
Insert a new cassette as described in:
9
Press <Yes> to return to the [Change Cassette] screen.
10
Enter the lot number of the cassette if it is different to the previous
cassette. This number is printed on the outer packaging box.
11
Press <Enter> to confirm the entry.
12
Press <New>. The cassette mechanism is set to the working position.
13
When the setting is complete, press <OK> to return to the [TS Cassette]
screen.
14
Press <Home> to return to the [Routine Monitor] screen.
The analyzer must be in the Stand-By status. If it is not, press <STOP> on the
[Routine Monitor] screen and wait for the Stand-By status to appear.
Step 6 to 9 of “To replace an empty URISYS 2400 cassette” on page C-18
When the new cassette is inserted, the [Routine Monitor] screen indicates the
number of remaining strips as “400” and the new loading date. The analyzer is now
ready for further processing.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......