Roche Diagnostics
Operator’s Manual · Version 1.0
D-9
URISYS 2400
7 General maintenance actions
Exchange user-replaceable parts
a
To disassemble the syringe
1
Loosen the screw of the syringe cover below the W/S syringes and remove
it.
2
Disconnect the tubing at the top of the syringe by turning the small,
colorless tube-retaining nut counterclockwise.
3
Place a dry absorbent gauze under the small, colorless tube-retaining nut
that is connected to the side of the lower blue part of the syringe and
disconnect this side-tube (some water will flow out of the side connection
of the syringe).
Do not open the syringe cover while the power switch is turned on. Otherwise your
fingers may be snagged with moving parts.
Figure D-1 Loosening top-tube retaining nut
Figure D-2 Disconnecting side tubing
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......