Roche Diagnostics
C-18
Operator’s Manual · Version 1.0
5 Daily operation
URISYS 2400
Replace an empty URISYS 2400 cassette
Replace an empty URISYS 2400 cassette
When a URISYS 2400 cassette is empty, an alarm message is generated and the
<Change TS Cassette> button on the [Routine Monitor] screen is enabled.
The status of the analyzer changes to S.Stop (TS Empty). The analyzer finishes
measuring the test strips being processed.
a
To replace an empty URISYS 2400 cassette
1
Press <Change TS Cassette> on the [Routine Monitor] screen. The [TS
Cassette (S.Stop)] screen appears.
2
Press <Remove>. The [Change Cassette] window appears.
3
Press <Execute>. The “Do not open the door” window message is
displayed, followed by an “Exchange cassette and press <Yes> key”
message.
4
Open the cassette compartment door.
5
Remove the cassette by pulling the handle on the left side of the cassette
and take out the cassette.
6
Open a new URISYS 2400 package, unfold the aluminum bag and cut it
open with scissors.
7
Remove the cassette from the packaging
and remove the two protection covers.
8
Place the cassette on the opened door with the label facing out and move it
smoothly into the cassette compartment.
9
Push the cassette completely to the rear of the compartment and close the
door firmly.
10
Press <Yes> to return to the [TS Cassette (S.Stop)] screen.
11
Enter the lot number of the cassette if it is different to the previous
cassette. This number is printed on the outer packaging box.
12
Press <Enter> to confirm the entry.
13
Press <Re-Start> to set the cassette mechanism to the working position.
After completion of the setting, the analyzer automatically resumes
Operation status and continues processing samples.
When the new cassette is inserted, the [Routine Monitor] screen indicates the
number of remaining strips as “400” and the new loading date.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......