Roche Diagnostics
B-22
Operator’s Manual · Version 1.0
3 Software
URISYS 2400
Installation
a
To change the range values
1
Press <Range Table> on the [Test Parameters] screen. The [Range Table]
screen is displayed.
2
Press <Test> to select the test parameter to be viewed. For each test
parameter, the range and reflectance values can be changed.
3
Press <Set Range> to change the range. The [Password] window appears.
4
Enter the 4-digit password.
5
Press <Enter>. When the correct password is entered, the [Set Range]
screen is displayed.
6
Press <Test> to select a test parameter.
7
Enter the range values. Press “<” or “>” to move through the different
fields.
8
Press <Enter> after each value is entered.
9
Repeat steps 6-8 if the ranges of other test parameters need to be changed.
10
Press <PgUp> to return to the [Range Table] screen.
a
Using arbitrary units
If arbitrary units (1+, 2+, 3+) should be used instead of concentration values
or in addition to concentration values, proceed as follows:
1
Call up the [Set Range] screen.
2
Press <Test> to select a test parameter.
3
Enter the arbitrary unit in the corresponding field and delete the
concentration value by pressing <Space> if you want to replace the
conventional or SI units.
4
Enter the arbitrary unit in the corresponding field without deleting the
concentration value if you want to used combined units, for example,
“2+ 25”
.
o
The screen allows for eight ranges to be set for each parameter. However, not all
of the ranges are used by the default settings.
o
If ranges are changed for conventional concentration units, these changes are
not transferred to the SI concentration units memory, and vice versa. To be
consistent, corresponding changes must be made in both range tables. (See
Table F-1 on page F-5.)
If combined units are used for the parameter Glucose, the entry of “4+1000” is not
possible because of the limitation of 5 digits per field. Use “4+999” instead.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......