a
To change the reflectance values
1
Press <Set Reflect>. on the [Range Table] screen. The [Password] window
appears.
2
Enter the 4-digit password.
3
Press <Enter>. When the correct password is entered, the [Set Reflectance]
screen is displayed.
4
Press <Test> to select a test parameter.
5
Enter the new reflectance values between the indicated values. Apply
moderate changes and validate the effect to the evaluation sensitivity.
6
Press <Enter> after each value is entered.
7
Repeat steps 4-6 if reflectance values of other test parameters need to be
changed.
8
Press <PgUp> to return to the [Range Table] screen.
a
To set default concentration ranges and reflectance values
1
Press <Default> on the [Range Table] screen. The [Password] window
appears.
2
Enter the 4-digit password.
3
Press <Enter>. When the correct password is entered, the [Default]
window appears to confirm the change to the default settings.
4
Press <OK> to confirm. The changed concentration ranges and reflectance
values are set to default.
5
Press <PgUp> to return to the [Test Parameters] screen.
o
The screen allows for eight values to be set for each parameter. However, not all
of the values are used by the default settings.
o
Reflectance values that have been changed will be marked with a “#” flag on
the [Range table] screen as well as in the Parameter Settings printout. This flag
will also appear with the corresponding test parameter on the result printout.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......