Roche Diagnostics
Operator’s Manual · Version 1.0
D-15
URISYS 2400
7 General maintenance actions
Exchange user-replaceable parts
9
Perform “Air Purge”, if necessary more than once, to fill the syringe with
water and to get rid of any remaining air bubbles.
Exchange mains fuses
a
To exchange mains fuses
1
Set the main switch to OFF position.
2
Remove the power cord from the power inlet socket of the instrument.
3
Remove the fuse socket located directly below the power inlet socket of the
instrument.
4
Exchange the fuse(s) as necessary.
5
Reinsert the fuse socket.
6
Reinsert the power cord into the power inlet socket of the instrument and
set the main switch to ON.
Figure D-12 Replacing the retaining nut
The top and side tubing retaining nuts can be damaged if excessive force is applied
when replacing them. Take care not to cross-thread these nuts.
Looseness of the retaining nut may result in incorrect data. Tighten the retaining nut
securely by hand while being careful about the packing orientation.
Use only fuses that are specified in this Operator Manual, (see “Technical
specifications” on page F-2). The use of other fuses is a breach of the warranty.
Summary of Contents for URISYS 2400
Page 1: ...Operator s Manual Version 1 0 URISYS 2400...
Page 4: ...Roche Diagnostics iv Operator s Manual Version 1 0 URISYS 2400...
Page 6: ...URISYS 2400 Roche Diagnostics vi Operator s Manual Version 1 0...
Page 14: ...Roche Diagnostics xiv Operator s Manual Version 1 0 URISYS 2400...
Page 15: ...Part A is an overview of the URISYS 2400 Overview A...
Page 16: ......
Page 23: ...Part B is the system description of the URISYS 2400 System description B...
Page 24: ......
Page 57: ...Part C contains the operating procedures for the URISYS 2400 Operation C...
Page 58: ......
Page 102: ...Roche Diagnostics C 44 Operator s Manual Version 1 0 6 Special operations URISYS 2400 Checks...
Page 103: ...Part D describes all maintenance procedures relating to the URISYS 2400 Maintenance D...
Page 104: ......
Page 122: ......
Page 147: ...Part F describes the technical specifications relating to the URISYS 2400 Appendix F...
Page 148: ......
Page 155: ...Glossary and index G...
Page 156: ......