P
AGE
49
D
IAGNOSTIC
F
UNCTION
C
HAPTER
6
Diagnosis in 15 Steps
START
After turning the switch S12 to the ON position, the test steps described
below are called up by pressing the START button
D
IAGNOSIS
1
C
HECK
LED
S
Press the START button briefly.
The LEDs light one after the other:
1.
G
REEN
A
NALYSIS GOOD
2.
R
ED
A
NALYSIS BAD
3.
Y
ELLOW
A
NALYSIS ACTIVE
4.
B
LUE
REAGENT
5.
R
ED
A
LARM
green
red
yellow
blue
red
D
IAGNOSIS
2
1.
X
START
C
HECK THE
P
USH
B
UTTONS
Press the START button briefly.
Pressing the following buttons will cause
the LEDs to light jointly with the green
LED.
FLUSH
GREEN
+
YELLOW
INDICATOR
GREEN
+
BLUE
RESET
GREEN
+
RED