C
HAPTER
5
M
AINTENANCE AND
S
ERVICE
P
AGE
45
M
AINTENANCE
S
TEP
3
FLUSH
INDICATOR
+
R
EFILL REAGENT AND
R
ESET THE
S
TORAGE
C
OUNTER
1. remove empty reagent bottle
→ open bottle cap by turning
→ remove the suction pipe carefully
→ if reagent fluid leaks, wipe up with a
paper towel
→ dispose of empty bottle
2. replace full reagent bottle
→ remove the cap
→ insert the suction pipe carefully
→ screw bottle cap firmly shut
→ pay attention to a correct positioning
of the cap and the adapter.
3. de-aerate the reagent pipe
→ fill the measuring-chamber with
water by pressing the FLUSH
button
→ dose reagent until the pipes and the
cartridge of the peristaltic pump are
completely filled with reagent and
no air bubbles are visible by
pressing INDICATOR button
4. confirm bottle change and reset counter
→ press INDICATOR and RESET
button simultaneously until the right
red and the blue LEDs light up
briefly
blau rot
NOTE
E