C
HAPTER
4
O
PERATION AND
U
SE
P
AGE
33
First result suppression REL 1
►
Relay function REL 1
Relay REL 1 signals that the limit has been exceeded. You can choose between an
impulse contact (3 und 60 seconds) and a permanent contact.
Possibilities for permanent contacts:
1. Further analyses will be carried out and if the sample is below the limit, REL 1 will
be deactivated automatically.
2. If the limit is exceeded, no further analyses will be carried out and REL 1 must be
deactivated manually (RESET). Only then can more analyses be carried out.
S3
Function
OFF
No first result
suppression
ON
First result suppressed
Function relay REL 1
S4
S5
Function
OFF
OFF
impulse contact
3 seconds
OFF
ON
impulse contact
60 seconds
ON
OFF
permanent contact
no analysis stop
ON
ON
permanent contact
analysis stop
ON