4.4 Setting the programme switches
The operating requirements of the Limit Analyser are programmed using the small DIP
switches S1
– S12.
Switch off the unit and open the lid of the controller
→ WARNING supply voltage 85 -264 VAC 47 – 63 Hz
→ The DIP-switches (dual in-line package) are located on the
reverse side of the display board. The display board is
screwed to the unit lid and should not be removed.
Also located on this board are the 2 16 position rotary switches
for setting the flush time and the delay times.
ON
1
2
3
4
5
6
7
9
8
10 11 12
OFF
Factory settings
Flush time
4 minutes
Position 6
4 min.
Delay time
Off
Position 0
0 min.
Analysis interval
30 minutes
S1 OFF
S2 ON
First result suppression
Yes
S3 ON
Function
Relay 1
Contact closed =
continuous analysis
S4 ON
S5 OFF
Limit value iron content
0.2 mg/l
S6 OFF
S7 ON
S8 ON
Monitoring
Signal when limit
exceeded
S9 OFF
Indicator type
Fe-LRS 0.03-0.6 mg/l
S11 OFF
operating mode
Analysis
S12 OFF
Work on electrical equipment of the plant / machinery must be performed
by a qualified electrician! Observe all applicable electrical regulations.
→ WARNING supply voltage 85 – 264 VAC 47 – 63 Hz
For the operation of the rotary switches and for setting the slide switch you
need a small insulated screwdriver
→
Please use only good quality tools, this will help avoid damage to
sensitive components
D
IP
S
CHALTER
WARNING
ELECTRICITY
NOTE