Gas Tracer Operator’s Manual, CSA Version
Using Maintenance Mode • 147
Updating the Datalog Overwrite Setting
With OVER WRT set to ON (factory setting), the Gas Tracer writes over the oldest
data with new data when the data logger memory is full.
With OVER WRT set to OFF, the Gas Tracer stops saving data to the data logger
when the data logger memory is full.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the OVER WRT menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
Updating the Calibration Interval
This CL INTVL setting defines the amount of time between calibrations. The
minimum setting is 1 day and the maximum setting is 365 days. The factory
setting is 90 days.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the CL INTVL menu item.
Press and release the POWER ENTER button. The screen will indicate the
current setting.
2. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
Tip:
Press and hold the AIR
▲
or (SHIFT)
▼
button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting and return to
the main menu.
On
MAINTENANCE
OVER WRT
90
MAINTENANCE
CL INTVL