154 • Using Maintenance Mode
Gas Tracer Operator’s Manual, CSA Version
Updating the Leak Check/Bar Hole Mode Setting
With L./B. MODE turned OFF, the Gas Tracer will automatically enter Normal
Mode when it is turned on.
With L./B. MODE set to bH, you will be prompted to choose between Normal
Mode and Bar Hole Mode when the Gas Tracer is turned on.
With L./B. MODE set to LC (factory setting), you will be prompted to choose
between Normal Mode and Leak Check Mode when the Gas Tracer is turned on.
With L./B. MODE set to LC bH, you will be prompted to choose between Normal
Mode, Bar Hole Mode, and Leak Check Mode when the Gas Tracer is turned on.
1. Use the AIR
▲
or (SHIFT)
▼
buttons to display the L./B. MODE menu item.
2. Press and release the POWER ENTER button. The screen will indicate the
current setting.
3. Use the AIR
▲
or (SHIFT)
▼
button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and return to
the main menu.
MAINTENANCE
L./B. MODE
OFF
MAINTENANCE
L./B. MODE