Gas Tracer Operator’s Manual, CSA Version
Display Mode • 47
3. Since the oxygen concentration in fresh air is above both oxygen alarm points,
the Gas Tracer will go into alarm if turned on in a fresh air environment.
Display Mode
This section describes using the Gas Tracer in Display Mode. With the Gas Tracer
in Display Mode, you can:
• set the combustible gas channel range
• display peak readings
• display STEL and TWA readings (CO only)
• display full scale, warning, alarm, STEL (CO only), and TWA (CO only) values
• display time in operation
• display the date and time
• display remaining log time
• clear the data logger
• turn the pump on or off
• select a user ID
• select a station ID
• display snap logging data
• turn peak bar function on or off
To enter Display Mode, press and release the DISPLAY (ADJ) button while in
Measuring Mode. To scroll from one screen to the next press and release the
DISPLAY (ADJ) button.
NOTE:
Each screen displays for 20 seconds. If you do not press the DISPLAY
(ADJ) button to scroll to the next screen or press the POWER ENTER
button to enter an item within 20 seconds, the Gas Tracer automatically
returns to Measuring Mode.