37 • Calibrating Instruments with Standard Sensors
To perform a calibration on an instrument:
1. Confirm that the AC Adapter is connected to the SDM-E2 and to an AC wall
socket.
2. Press and hold the SDM-E2’s POWER button. The LEDs will turn amber.
3. When the BUMP
and CAL
LEDs turn off, release the POWER button.
4. The COPY LED will be off or on steadily and the CHARGE LED will be blinking
green if the SDM-E2 is operating properly or solid red if there is a system
failure. The amount of free memory in the SDM-E2 will dictate the condition of
the COPY LED (see “Available Memory in the SDM-E2” on page 69).
5. Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of
the SDM-E2 and slide the instrument forward until it falls into place as
illustrated below.
Figure 12: Inserting the EAGLE 2
WARNING:
Inserting the EAGLE 2 improperly may damage the exhaust bellow at the
back of the SDM-E2.