35 • Bump Testing Instruments with Standard Sensors
The current gas readings will be displayed along with the bump test
results. A “C” will appear next to the channels indicating they are
being calibrated. The BUMP
LED will continue to flash amber and
the CAL
LED will begin flashing amber to indicate a calibration is
taking place.
b. At the end of the calibration, the instrument displays the calibration
gas readings at the end of the gas application and the results from
both the bump test and the calibration with a P (pass) or an F (fail).
c. The BUMP
LED and CAL
LED will continue to flash amber
while the docking station performs a fresh air purge for the time
defined by the AIR TIME bump test parameter. After the purge is
complete, the BUMP
LED will turn solid red. If all channels
passed the calibration, the CAL
LED will turn solid green. If any
channels failed the calibration, the CAL
LED will turn solid red.
18. After a successful or failed bump test,
•
To perform any other operations:
To perform another bump test, press and hold the BUMP
button until the
pump starts. To perform a calibration, press and hold the CAL
button
until the pump starts. To return to the home screen, press and hold the
EDIT ENTER button until the home screen appears.
•
To turn the EAGLE 2 off:
If the bump test was successful, the instrument will shut off after 15
CH4 29%LEL BF C
OXY 18.0vol% BF C
H2S 25.0ppm BP C
CO 45ppm BF C
---
---
CH4 29%LEL BF CF
OXY 18.0vol% BF CF
H2S 25.0ppm BP CP
CO 45ppm BF CP
---
---