8. Alarm Function
8-2. Fault alarm activation
- 68 -
8-2. Fault alarm activation
"Fault alarm" is triggered using a buzzer sound and blinking alarm lamp when an abnormality is detected
in the gas detector. (Self-latching)
When an alarm is triggered, one of the following fault details is displayed on the LCD.
・
System abnormalities:
SYS
□□□
・
Sensor abnormalities:
SENSOR
・
Calibration abnormalities: AIR CAL
・
Battery voltage low:
BATTERY
・
Low flow rate:
LOW FLOW
・
Clock abnormalities:
CLOCK
Alarm lamp Repeatedly blinks at about one-
second intervals.
Buzzer
Repeatedly sounds intermittent
beeps at about one-second
intervals.
LCD display
Display example of low flow rate
(LOW FLOW)
If a fault alarm is triggered, determine the cause and take appropriate action.
If the gas detector has problems and is repeatedly malfunctioning, contact RIKEN KEIKI immediately.
NOTE
•
A low flow rate alarm (FAIL LOW FLOW) can be reset by pressing the RESET button. For other
fault alarms, turn off the power and then promptly contact RIKEN KEIKI.
Note that a clock abnormality alarm (FAIL CLOCK) can be reset by pressing the RESET button, but
the internal clock is not operating normally and thus the data logger cannot be recorded normally.
•
For details about malfunctions (error messages), see "Troubleshooting" (P. 75).