3. Product Components
3-2. Names and functions for each part
- 12 -
<LCD Display>
No.
Name
Function
(1)
Pilot indicator
Displays the operating status in the detection mode. Normal:
Blinking
(2)
Flow check display
Displays the drawing status. Normal: Rotating
(3)
Range mode display
Displays the Lo/Auto range with an icon. No icon is displayed in the
Hi range mode.
(4)
Battery level icon
Displays a reference of the battery level.
(5)
Gas name display
Displays detected gases name.
(6)
Concentration and unit
display
Displays gas concentration and unit.
(7)
Full scale and bar
display
Displays the level of gas concentration with the bar meter
as well as the full scale value.
(8)
Clock display
Displays the current time.
NOTE
•
The meanings of battery level icons are as follows:
: Sufficient/ : Low/ : Needs charging
If the battery level further drops, the inside of the battery icon starts to blink ( ).
•
Range mode display
・
Lo: Fixed to the low range (0 - 10.0 vol%)
・
Auto: Automatic range switching that switches between the low and high ranges automatically
・
Hi: Fixed to the high range (0 - 100 vol%)
(8) Clock display
(7) Full scale
and bar display
(6) Concentration
and unit display
(5) Gas
name
(1) Pilot indicator
(2) Flow check
display
(3) Range mode
display
(4) Battery level
icon