5. How to Set Display Mode
5-2. Concentration displayed gas reading setting (base)
- 32 -
3
Select a gas name
to read with the
▼/▲
button and then
press the ENTER
button.
▲
▼
▼
* The gas reading setting will be
cleared at power-off.
After "END" is displayed,
the gas detector returns to
the display mode menu.
▼
NOTE
•
The reading accuracy on the list of specifications is applied to the calibration gas (CH4) only.
•
The concentration displayed for a converted reading is a reference value. To display an accurate
concentration, calibration using the gas to be measured is required. Therefore, request RIKEN
KEIKI for span adjustment using the gas to be measured.
•
For a list of the gases that can be read, see "NP-1000 gas list" in the following page.
CAUTION
•
To perform the concentration displayed gas reading setting, see "NP-1000 gas list" in the
following page.