. Calibration
7-9. Bump test condition setting
- 63 -
7
Press the
▲
button.
8
Set whether or not to
perform span
calibration after "F"
(Failure) is displayed
as a result of
diagnosis.
Press the ENTER button.
When the display starts
blinking, select a value with
the
▲/▼
button.
When the ENTER button is
pressed, “END” is displayed
and the setting is confirmed.
* Set range
on: Span calibration
performed
off: Span calibration not
performed
9
Press the
▲
button.
10 Press the ENTER
button with “ESCAPE”
displayed.
The gas detector returns to
the calibration mode menu.
(Buzzer sound: Once <blip>)