Phoenix HT User Manual
© 2007–2012 Rigaku Automation, Inc.
38
401718 Rev A
5
Adjust the jog increment to the desired resolution using the pulldown box at the bottom
of the page, then jog the X and Y axes the distance estimated in step 3.
Keep in mind that the shelves will need to move in the opposite direction.
6
Click
Set Taught Position
, then
Stop Teaching
then
Save and Close
.
5. Test dispensing:
1
Reconnect to the instrument and create another test plate.
2
If the drops are still not merging, determine the distance you need to move the
nanodispenser tip, and then repeat the substeps under step 4.
3
Ensure that you keep track of the net distance you have moved from your original starting
position because you will need to apply this to the rest of the dispense positions.
6. Apply offsets to all positions:
1
Once you are satisfied with the drop placement on one position, the position shift must be
applied to all of the other nanodispense positions.
2
Open the calibrate stage view, connect, and then select the nanodispenser in position 2.
3
Click
Go above XY position
and then jog the axes by the net distance you determined for the
first plate position.
4
Click
Set Taught Position
, then
Stop Teaching
, then
Save
to save the adjustments made.
5
Repeat substeps 1 through 3 under step 6 for all positions into which the nanodispenser
can dispense (Positions 1-6).
6
Once you have made all the necessary dispense adjustments, click
Save and Close
to exit
the calibrate page.
7. Run a test plate on each of the positions using
FAT Dispense 100nl+ 100nl on Lucite Plate.pro
to
ensure that you are satisfied with the drop placement.