Phoenix HT User Manual
© 2007–2012 Rigaku Automation, Inc.
35
401718 Rev A
7. Observe that the vacuum pump runs for a short time, then stops.
If it runs continuously, remove and reinstall the new tip.
If the vacuum pump is working properly, obtain one sample tube filled with water and
gently submerge the nano-tip in it. If the tip bubbles the liquid, replace the nano-tip before
going further.
8. Calibrate the nano-tip for all nano-accessible tray positions.
1
Select
Disconnect
to disconnect the instrument.
2
Remove any obstacles or consumables on the tray positions.
3
Ensure that the predrilled holes for the accessible tray positions are completely visible.
Using a flashlight is highly recommended.
9. If the previous nano-tip was well-aligned, proceed to “Aligning the Nano Dispenser” on
page 37 to ensure that the new nano-tip is properly configured.
If the previous nano-tip was not well aligned or the alignment is uncertain, proceed to
“Calibrating the Nano Dispenser” on page 35, then complete the process with “Aligning the
Nano Dispenser” on page 37.
5.5. Calibrating the Nano Dispenser
1. In the ARI Phoenix application, select
Calibrate Stage
in the
Configuration
menu.
Figure 4. Calibrate Stage Step
Connect button
Calibrate Stage selection