Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and landmobile radios, amateur radio, AM and FM radio bro-
adcast and TV broadcast cannot be predicted theoreticallywith accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic
sitesurvey should be considered. If the measured field strength in the location in
which the
ri-
thermo
®
fastPRObe Thermometer is used exceeds the applicable RF
compliance level above, the
ri-
thermo
®
fastPRObe Thermometer should be ob-
served to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the
ri-
thermo
®
fastPRObe Thermometer.
b.) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the
ri-
thermo
®
fastPRObe Thermometer.
The
ri-
thermo
®
fastPRObe Thermometer is intended for use in an electro-
magnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the
ri-
thermo
®
fastPRObe Thermometer can help
prevent electromagnetic interference by maintaining a minimum distance bet-
ween portable and mobile RF communications equipment (transmitters) and
the
ri-
thermo
®
fastPRObe Thermometer as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
(W)
Separation distance according to frequency of trans-
mitter
(m)
150 kHz to 80
MHz
d =12√P
80 MHz to 800
MHz
d =1,2√P
800 MHz to
2,7 GHz
d =2,3√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recom-
mended separation distance d inmetres (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is themaximum output pow-
er rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propaga-
tion is affected by absorption andreflection from structures, objects and people.
Summary of Contents for ri-thermo fastPRObe
Page 53: ...53...
Page 79: ...79...
Page 131: ...131 1 1 1 ri thermo fastPRObe Riester 93 42 93 42 EEC Ries ter 1 2 MD B IPX1 IP DEHP...
Page 132: ...132 Da consumarsi entro YYYY MM anno Mese REF UDI C F 2002 96 EC Recall 0124...
Page 135: ...135...
Page 139: ...139 4 4 1 4 1 1 1 1 2 2 1 3 1 4 1 5 6 7 1 2...
Page 140: ...140 4 2 8 60...
Page 141: ...141 4 3 4 3 1 4 3 2 4 3 3 4 3 4 4 3 5 4 3 6 4 3 7 4 3 8 4 1 1 2 1 3 1 2...
Page 142: ...142 ri thermo fastPRObe ri thermo fastPRObe 4 4 1 1 2 1 3 1 2 3...
Page 143: ...143 4 1 2 5 1 4 5 4 5 1 4 5 2 33 5 C 92 3 F 4 5 3 4 5 4 4 5 5 4 5 6 4 5 7 4 5 8...
Page 149: ...149 9 IEC 60601 1 2 IEC 60601 1 2 9 1 9 1 1 9 1 2 9 1 3 9 1 4 9 1 5 60 60 60 12 12 12...
Page 150: ...150 IEC 60601 1...