10 - ERROR MESSAGE / CORRECTION
52
timeout printer internal
the internal printer is temporarily discon-
nected
53
Set of data points is missing
Check multi-standard functions
54
nonlinear measurement: value out of range
check the absorbance of the sample
55
number of given data points < 2
add data points
59
error in the automatic measuring operation (remote)
check interface
60
error at multiplexer of operational amplifier
contact service partner
62
free method number not found
check method memory
63
wrong address at multi-standard loading
check multi-standards
64
current method not found in monthly memory
check QC data of method
65
more than 50 QC methods defined
delete unused QC methods
66
internal clock is off. QC data not storable
switch on internal clock
67
BCC error in dataset of QC method values
check current method
68
free monthly memory not found at QC
delete unused QC methods
69
free space in daily memory not found at QC
empty QC daily memory
70
error at QC calculation
check QC data
71
QC serum not found
check QC data
72
wrong address at saving test results
send results through serial interface and
delete results (chapter 8.2.8)
73
results memory full, it will be overwritten
overwrite results memory or send and
delete results (chapter 8.2.8)
74
BCC error at sending result string
send results through serial interface and
delete results (chapter 8.2.8)