Calculation procedure .................................................... CP 16
Characteristic ...................................................... DELTA R1R2
Method ..................... Difference measurement of two reagents
Calculation formula ...................................................... C =
A
S
Factor ............................................................... given / entering
Time / Delta T1 ........................................ entering (20 to 255s)
Measure time R1 T2 ................................ entering (0 to 1800s)
Measure time R2 T3 ................................ entering (0 to 1800s)
Delay T4 .................................................. entering (0 to 1800s)
PHOTOMETER 4040+ # 1001
V2.2a dd/mm/yy D
LAB.: RIELE BERLIN
USER 1: M.MUSTERMANN
DATE: 07/17/17
TIME: 08:44:12
METHOD 20: DELTA R1R2
PROGRAM: 16
FACTOR: 1.000
WAVELENGTH: 546nm
TEMPERATURE: 37C
TIME/DELTA: 30s
DELAY: 0s
TIME REA.#1 130s
TIME REA.#2 130s
UNIT:
- - - - - - - - - - - -
MEASURE ZERO
E1 1 0.285 [A]
E1 2 0.285 [A]
E1 3 0.285 [A]
E2 1 0.165 [A]
E2 2 0.165 [A]
E2 3 0.165 [A]
NO. RESULT
1
-0.116
2
-0.116
3
-0.116
Start method selection in the main menu.
See chapter:
4.1 - Measurement with programmed methods
4.2 - Measurement with basic methods
In the case of activated printer the print out of the method data
follows.
The measuring window is shown.
Insert / measure zero solution
Insert / measure samples with reagent 1
Insert / measure samples with reagent 2
The user is guided through the measuring procedure on the
screen.
The result will be displayed