4040+e_22.docx / 06.10.17
ROBERT RIELE GmbH & Co KG
Page 65
10 - ERROR MESSAGE / CORRECTION
9
check sum error in the data record of the device basic
setting (status, ADC correction)
automatic error remedy
10
division by a too small value (< 0.001 A)
check filter; check standard;
check measuring solution
11
invalid calibration curve
Select valid number
12
setting to zero not possible (zero value is < 32768 cycle)
check LED; check filter;
check zero solution
13
setting to zero not possible (zero value is > 983039 cycle)
repeat optic adjustment (chapter 8.2.1);
check LED; check filter;
check zero solution
14
invalid standard
measure valid standard solution
15
no parameter memory vacant (too little memory for non-
linear methods)
delete a no longer actual nonlinear
method
16
method no. is occupied
select other method no.; delete a no
longer actual nonlinear method
17
check sum error in the parameter memory (nonlinear
method)
program method newly
18
at calculation overflows in nonlinear method
check factor; check parameter
19
clock malfunction
20
overflow at Kinetic
check measuring solution
21
overflow at Kinetic
check measuring solution
22
overflow at Kinetic
check measuring solution
23
overflow at Kinetic
check measuring solution
24
overflow at Kinetic
check measuring solution
25
overflow at Kinetic
check measuring solution
26
overflow at Kinetic
check measuring solution
27
overflow at Kinetic
check measuring solution
28
overflow at Kinetic
check measuring solution
29
wrong input of deltas or time per delta
restart method
30
battery empty
contact service partner
31
communication: wrong data format
contact service partner
32
communication: sent data not plausible / not interpretable
contact service partner
33
communication: mentioned module does not answer in a
certain time
check interconnecting cable;
check mentioned module
34
communication: overflow send/receive buffer
reduce amount of data at the
communication partner
35
remote control: wrong method number
external software problem
36
remote control: unknown command
external software problem
37
remote control: wrong data format
external software problem
38
check sum of operating system bank 0 damaged
contact service partner
39
check sum of operating system bank 1 damaged
contact service partner
40
timeout at reception from module 2
switch off/on device
41
check sum error at reception from module 2
switch off/on device
42
NAK at reception from module 2
switch off/on device
46
filter position out of tolerance
contact service partner